5 calibration procedures for A&D indicators, balances, and scales, from our technicians' notes.
Turn off Turn throw switch in back press zero CAL1 press stby oper and you should get ................. in display enter the calibration weight using the keys place calibration weight on scale and press stby oper you should get.................. slide switch back take off calibration weight You be done
OPEN BACK OF SCALE PUSH SWITCH TO "ON" PRESS "UNITS" DISPLAY WILL READ "CAL O" PRESS "ZERO" PRESS UNITS DISPLAY READS "CAL 3" WAIT FOR CIRCLE IN UPPER LEFT CORNOR PRESS "ZERO" DISPLAY WILL READ "END" SLIDE SWITCH BACK
(FV-30K 60LB/40LB)(FW-10K) CAL 1: MAX CAP. IN KG = 150/60/30KG. CAL 2: 2/3 CAP. IN KG = 100/40/20KG. CAL 3: MAX CAP. IN LB = 300/120/60LBS. CAL 4: 2/3 CAP. IN LB = 200/80/40LBS. OPEN UP BATTERY COMPARTMENT REMOVE SCREW FLIP CAL SWITCH TO ON IT WILL READ 9.798 PRESS "MODE/UNITS" "CAL O" STABLE LIGHT PRESS "REZERO" "CAL1" PRESS "MODE' TO SELECT SETTING CAL 1,2,3,4 (HIGH/LOW/GO WILL ROTATE SETTINGS) PUT calibration weight ON (200LBS.) AND WAIT FOR STABLE LIGHT PRESS "REZERO" "CAL END" IS DISPLAYED PUT SWITCH BACK
HOLD SAMPLE AND PRINT CAL OUT CAL 0 PRESS SAMPLE SHOWS Cal press print cal 0 prees print add calibration weights press print end remove calibration weight done
A&D EP 60 KB Throw switch on side cal 0 press print Enter Cal Wt press tare (possibly twice) calF put on calibration weight press tare Cal End