1 calibration procedure for Adam Equipment indicators, balances, and scales, from our technicians' notes.
Turn the Power OffTurn the power back on , during the counting from 9 to 0 press func/c key. The display will show "CAL" for a few seconds. While it is showing "CAL" press the SMPL,PRINT and TARE keys in sequence to enter the calibration section. The display will show "unload". Remove any calibration weight from the platform. Press ZERO/ENTER key. The display will show the last calibration weight used. If this is correct you cn continue by pressing the ZERO/ENTER key. If it is not correct use the arrow keys to change the calibration weight value. When it is correct press the ZERO/ENTER key. If the calibration is acceptable the displaywill show "PASS" and then retur to normal. If an error message is shown try calibration again as a disturbance may have prevented a successful calibration.