4 calibration procedures for Digi indicators, balances, and scales, from our technicians' notes.
select scale to calibrate using" scale" button Press and hold the rezero key and press 8715 let go of rezero empty scale press rezero put calibration weight on scale enter amount of cal wt press "*" press mode twice do this per scale attached
Press mode repeatedly until maintenance. Press and hold rezero then press 8715. release rezero press tare (right arrow key) till span adjust press enter the left of the display is the scale number to change this press + to the right of that is the scale capacity. To the riqht of that is the calibration weight. The calibration weight has to be entered with the same amount of decimals as the scale capacity. To change either number use the tare to highiite then enter new ualue. Once all values here are ok press enter. Empty scale push rezero add calibration weight push enter * press mode till operation press enter
For the 20 the Jumper JP-2 must be enabled on the main board Press And hold the rezero button and press the up button to the right of the tare button then tare twice It then displays the scale increments change with arrow keys or accept and push rezero Enter the capacity of the scale push rezero Enter the calibration weight you will use to cal push rezero cal 0 appears push rezero with nothing on scale cal sp appears add calibration weight and press rezero then press tare to get out
Remove label casette In the back of opening there is a 3 position switch. Move this switch into the middle position. Select scale to calibrate using scale button. Turn key to program Press and hold rezero while pressing 8715 release rezero. Empty scale push rezero add calibration weight enter calibration weight push enter turn key back to operate move switch back to original position