Calibration procedures

How to calibrate a Vibra Screw scale.

1 calibration procedure for Vibra Screw indicators, balances, and scales, from our technicians' notes.

Reference only. Every digital scale has a different sequence of buttons to calibrate it, and most need a known calibration weight — typically calibrating at both zero (dead load) and span (the known-weight portion). Have a known weight ready before you start. These instructions are provided for reference only and Progressive Scale takes no responsibility for their use. If you need an on-site scale calibration in the CT, NY, or MA area, call 203-792-2854.
Filling System

Vibra Screw Filling System

Throw cal switch on dip switches located at the bottom of aluminum box. Press mode to get to gross mode press the 8 button (also says cal). Displays SPCd press enter enter 5321 then enter setup press enter till cal press enter Hd Cal Press enter "Zero" displays press enter "-" displays press test clr 0.0 displays press enter span displays place calibration weight on scale press enter "-" displays press test/clr enter calibration weight press enter lincor press enter 0.0 displays press enter return displays press enter end cal displays press enter switch back