2 calibration procedures for Filing indicators, balances, and scales, from our technicians' notes.
Filing DR2100A 9090 ENTER DEL TOGGLES Y N MENU ENTER FOLLOW INST IF CAL SW OFF OPEN BOX AND THROW SWITCH
Filing DR-5000 The F1 & F2 buttons are shut off points for filling applications Cal Procedure To get into calibration mode press "9","9","Re-Zero" This calls the true zero. Go to the pot labeled R-40 and get as close as possibleto zero Push zero button Apply load Go to R-41 and dial to correct calibration weight Remove calibration weight Press "Re-zero" Re adjust if needed Once that is accomplished Call up true zero again. Go to R-40 and turn to negative 1/2 the scale capacity (-2500 to -2600lb) then rezero with button re apply calibration weight if incorrect, turn fine span pot 1/4 clockwise for up or 1/4 counter clock wise for down Do not watch the indcator because this will screw you up. Re zero the scale and re apply the calibration weight. repeat the procedure until it is correct.