Calibration procedures

How to calibrate a Acculab scale.

3 calibration procedures for Acculab indicators, balances, and scales, from our technicians' notes.

Reference only. Every digital scale has a different sequence of buttons to calibrate it, and most need a known calibration weight — typically calibrating at both zero (dead load) and span (the known-weight portion). Have a known weight ready before you start. These instructions are provided for reference only and Progressive Scale takes no responsibility for their use. If you need an on-site scale calibration in the CT, NY, or MA area, call 203-792-2854.
Balance

Acculab Balance

Empty Scale Press and hold Cal/mode until calibration weight appears in display Put calibration weight on scale Wait until a beep. Tthe cal calibration weight will blink and it will go back into weigh mode

Balance

Acculab S50 AND S100

PRESS MODE TO "KG" MODE PRESS AND HOLD "CAL" UNTIL CAL. WT. MODE APPEARS PRESS "TARE" (WHILE CAL IS STILL FLASHING) "CAL O" WILL BE DISPLAYED PAUSE 2 SECONDS PLACE CORRECT CAL. calibration weight ON SCALE "CAL F" DONE WHEN DISPLAY READS CORRECT calibration weight PRESS "TARE" TO RESET ZERO

Balance

Acculab VA8000

Hold F1 & on /off self test -c- press tare f1 cal displayed put on calibration weight 5000g turn off