Calibration procedures

How to calibrate a Sterling scale.

1 calibration procedure for Sterling indicators, balances, and scales, from our technicians' notes.

Reference only. Every digital scale has a different sequence of buttons to calibrate it, and most need a known calibration weight — typically calibrating at both zero (dead load) and span (the known-weight portion). Have a known weight ready before you start. These instructions are provided for reference only and Progressive Scale takes no responsibility for their use. If you need an on-site scale calibration in the CT, NY, or MA area, call 203-792-2854.
Counting Scale

Sterling Sterling

Sterling 825 Counting Scale 1)Press Scale Select to go to the scale you are calibrating 2) Press and hold Gross/Net Key until Params 3) Now the NTEP parameter along with its value is displayed 4)Press the print key or zero key till FCAL 5)Select FCALT with tare key and press Print 6) Shows current number f scales to change tis pram press tare or press print to move on 7) This is capacity change it or press print to move on 8)Graduations press print or change 9)Shows Cal Units print to move on 10) CAL F Change this to CAL T using Tare button press print to begin cal a)Clear scale and press print b)Step A8 press print c)Place calibration weight on scale, enter calibration weight value and press print 11)SPN 0.0 this is the span error press print 12)Shows caacity of next scale. Repeat steps 7-11 13)Gross/Net or reset key will get you out at the end