Calibration procedures

How to calibrate a GSE scale.

4 calibration procedures for GSE indicators, balances, and scales, from our technicians' notes.

Reference only. Every digital scale has a different sequence of buttons to calibrate it, and most need a known calibration weight — typically calibrating at both zero (dead load) and span (the known-weight portion). Have a known weight ready before you start. These instructions are provided for reference only and Progressive Scale takes no responsibility for their use. If you need an on-site scale calibration in the CT, NY, or MA area, call 203-792-2854.
Scale

GSE -350

GSE 350 Indicator Press zero and select simultaneously press zero,print,units and tare one after the other scale should say first zero. Press tare with nothing on the scale use th print and units keys to enter the load calibration weight press tare Add calibration weight press tare press tare till out

Scale

GSE 450 GSE 460

GSE 450 & 460 Indicator 1) Press [Zero]+[Select] Simultaneously 2) Press [Zero] [PRINT] [UNITS] [ENTER] In that order 3) Remove any extraneous Load 4) Press[Select] until "New Zero" is dislayed 5) Press[Enter] 6) place cal calibration weight on scale 7) Key in the calibration weight value 8) Press [Enter] "Cal OK" 9) Press [Enter] 3 times

Scale

GSE 455 GSE 465 GSE 550

GSE 455-465-550-574-650-660 1) From the calibration weight screen keyy in 100 [Select] DIsplays Key in Code: The access code must be keyed in within 5 seconds 2) Press 54321 [ID] [Enter] 3) Press Enter 4) Apply cal calibration weight 5) Key in calibration weight and press enter 6) Cal OK Press [Enter] 7) Enter=Save press [Enter] 8) Enter=Exit [Enter]

Scale

GSE 625 622

GSE 625 or 622 Counting Scale Use the scale select key to choose the scale you wish to calibrate Press the [GSE][MODE] to put the instrument into offset display mode press the [GSE][ZERO]. The scale is now in the auto cal mode Before going any further enter the program jumper bypass code [23640][GSE][.] The lower display will prompt you through the set up of Auto Cal. First of all remove all calibration weight and press [Zero]. The scale will average the zero while it counts to 10 in the lower display. Counter will reseet when motion is sensed When tho count reaches 10 place the calibration weight on scale,enter that amount and press[ENTER] When the procedure is complete press[MODE] to return to offset displayed If you want to cal another scale use scale select and start all over or press [ZERO] to re-enter the Weigh Mode

Other manufacturers

More calibration procedures.