Calibration procedures

How to calibrate a Doran scale.

4 calibration procedures for Doran indicators, balances, and scales, from our technicians' notes.

Reference only. Every digital scale has a different sequence of buttons to calibrate it, and most need a known calibration weight — typically calibrating at both zero (dead load) and span (the known-weight portion). Have a known weight ready before you start. These instructions are provided for reference only and Progressive Scale takes no responsibility for their use. If you need an on-site scale calibration in the CT, NY, or MA area, call 203-792-2854.
Bench Scale

Doran 4200

remove small plate in back of scale. thow red switch press gross net c000.00 press zero press units add calibration weight use arrows to correct units to save gross net

Bench Scale

Doran 7000 DORAN

OPEN UP SQUARE IN BACK (RECTANGULAR) "FLIP RED SWITCH" "CO" PRESS "REZERO" (COUNTS DOWN) C=calibration weight VALUE PUT THAT AMOUNT ON PRESS "REZERO" (COUNTS DOWN) C=calibration weight VALUE PUT THAT calibration weight ON PRESS "REZERO" (COUNTS DOWN) DONE FLIP SWITCH BACK

Bench Scale

Doran 8000

Doran 8000 Remove back cover plate Throw red switch inside Displays AUGAI Press "conv" Display ENT 0 Put calibration weight on scale zero=up tare = down to correct calibration weight Press "conv" displays ent sp Take calibration weight off flip switch back

Bench Scale

Doran 8000XL

Doran 8000 xl press cal sw if cap ok press units cal 0 press zero put calibration weight on scale press units press print hold print to go lower or capicity to go higher till cal calibration weight is matched press zero done